Dosimetric Properties of the Newly Developed LiF:Mg,Cu,Si TL Material
DOI:
https://doi.org/10.3329/jsr.v5i1.11935Keywords:
Dosimetry, TL, TLD, DoseAbstract
The dosimetric properties of the newly developed thermoluminescence (TL) material (LiF:Mg,Cu,Si) at Korea Atomic Energy Research Institute (KAERI) were investigated. The energy response of the detector was performed for photon energies from 20 to 662 keV. The dose response for this TL material (LiF:Mg,Cu,Si) was linear up to 10 Gy and a sub-linear response was observed for higher doses. The reusability of this newly developed TL detector sufficiently satisfied IEC standards. Detection threshold of LiF:Mg,Cu,Si TL material was investigated and found to be 930 nGy by Harshaw 4500 TLD reader.
© 2013 JSR Publications. ISSN: 2070-0237 (Print); 2070-0245 (Online). All rights reserved.
doi: http://dx.doi.org/10.3329/jsr.v5i1.11935 J. Sci. Res. 5 (1), 25-31 (2013)
Downloads
127
179
Downloads
Published
How to Cite
Issue
Section
License
© Journal of Scientific Research
Articles published in the "Journal of Scientific Research" are Open Access articles under a Creative Commons Attribution-ShareAlike 4.0 International license (CC BY-SA 4.0). This license permits use, distribution and reproduction in any medium, provided the original work is properly cited and initial publication in this journal. In addition to that, users must provide a link to the license, indicate if changes are made and distribute using the same license as original if the original content has been remixed, transformed or built upon.